Fluorad 150
 
 
Testphantom for acceptance testing of X-ray exposure and fluoroscopy equipment (acc. to DIN 6868/150).

With the
"Fluorad 150"
testphantom it is possible to measure the following parameters:

- Dynamic Range
- Contrast Resol.
- Spatial Resolution
- Beam Quality
- Homogeneity

Dimensions:
300mm*300mm*18.5mm

The copper plate within the details is bedded in acrylglass plates.
The acrylglass is white, so that the light beam can be better seen under spotlight conditions.
The structures are painted on booth sides of the testphantom.


Fluorad A+D
 
 
Testphantom for constancy tests of digital and analogue fluoroscopy equipment (according to DIN 6868/4:2007).

With the "Fluorad A+D" testphantom it is possible to
measure the following parameters:

- Dynamic Range
- Contrast Resolution
- Spatial Resolution
- Beam Quality
- Homogeneity

Dimensions:

"Fluorad 30 A+D":
300mm*300mm*18.5mm

"Fluorad 20 A+D":
200mm*200mm*18.5mm

The copper plate within the details is bedded in acrylglass plates. The acrylglass is white, so that the light beam can be better seen under spotlight conditions.


DSA Testphantom
 
 
A special DSA testphantom system is needed in order to conduct the prescribed acceptance- and constancy-testing of the DSA-specific key parameters of Digital Subtractionangiography equipment in accordance with
DIN 6868 Part 150 and Part 4.

The test phantom contains a movable insert with aluminium strips for blood vessel simulation.
This insert is moved via a pneumatic system during the exposure series.


Fluoro Phantom "D"
 
 
To check the digital fluoroscopic techniques.












Sensor in absorption filter
 
 
Fits the "ALPHA plus" Dosimeter for the constancy test of all analogue and digital fluoroscopic equipment.

The sensor is located above the patient-equivalent-filter (25mm Al) under the insertion plate.

Energy range:
50kV - 155kV

Effective dose range:
0,1mGy/s - 500mGy/s